Table 5-3: 12 Point 2D Calibration Pattern
Footnote:
1. For best results, the tilt experienced during calibration should match that experienced
in service. For example, if the TRAX is restrained to a level plane in service, then
calibration should be in a plane, where “max. positive” and “max. negative” are 0°.
5.1.3 Limited-Tilt Calibration
A Limited-Tilt Calibration is recommended when 45° of tilt isn’t feasible, but >5° of tilt
is possible. It provides both hard-iron and soft-iron distortion correction. The
recommended 12 point calibration pattern given below is a series of 3 circles of evenly
spaced points, with as much tilt variation as expected during use.
Table 5-4: 12 Point Limited-Tilt Calibration Pattern