PNI Sensor Corporation DOC#1016505 r06
TRAX User Manual Page 16
Table 5-2: 12 Point Full-Range Calibration Pattern
Footnote:
1. Yaw listings are not absolute heading directions but rather relative heading
referenced to the first sample.
5.1.2 2D Calibration
A 2D Calibration is intended for very low tilt operation (<5°) where calibrating the
TRAX with greater tilt is not practical.
This procedure calibrates for hard and soft-iron effects in only two dimensions, and in
general is effective for operation and calibration in the tilt range of -5° to +5°. The
recommended 12 point calibration pattern is a circle of evenly spaced points, as given in
Table 5-3.